Spectrometer | RP Series Laser Analysis and Design Software

Author:


Generally speaking,SpectrumAn instrument is an instrument used to study the wavelength-related characteristics of light, matter or objects; its use is quite extensive: 

A spectrometer is an instrument that can spatially separate the spectral components of light and analyze the spectral components separately-for example, using photographic negatives or external photodetectors. The spectrocolorimeter used is usually a diffraction grating or a prism.
The spectrometer usually also contains some photodetectors for analyzing the light intensity. A spectrometer containing a large detector array can be used to record the spectrum of the light source without scanning in the raster direction. When equipped with intensity calibration, such devices are more specifically called spectroradiometers.
Other optical spectrometers are used to analyze the spectral properties of substances or objects, such as wavelength-dependent transmittance or reflectance. They are more specifically called spectrophotometers and find application in fields such as chemistry. Particularly high spectral resolution and high sensitivity can be obtained using a laser spectrometer containing some narrow linewidth tunable lasers. However, these can usually only cover a rather limited spectral region.

Optical andphotonA variety of spectrometers outside the field of science, such as devices used to measure particle velocity or particle size distribution. However, this paper focuses entirely on spectrometers that perform spectral analysis of light. When interested in the analysis of substances or objects, refer to the article on spectrophotometers. 
Measurements made with spectrometers typically provide the optical power spectral density (PSD) of light as a function of wavelength or frequency. Not all spectrometers provide a calibrated PSD; typically, the intensity readings are not calibrated and may be strongly dependent on the calibration factor (responsivity) for wavelength. 
There are also spectral phase interferometry methods that measure not only the power spectral density, but also the spectral phase. 
Some spectrometers also haveImagingfunction, called an imaging spectrometer. See articles on hyperspectral imaging and multispectral imaging. 
If only the spectral linewidth of the laser beam needs to be measured, and the detailed spectral shape does not need to be measured, other methods can be used, such as self-heterodyne linewidth measurement. With this method, one can measure very small linewidths, which are much lower than the resolution of typical spectrometers. 

Types of spectrometers

Spectrometers based on diffraction gratings or prisms 
 
 
Most spectrometers are based on some kind of polychromator, a device that can spatially separate different wavelength components of light. Typically, they utilize wavelength dependent diffraction on one or more diffraction gratings or wavelength dependent refraction on one or more prisms. 
Typically, the incident beam is collimated (made parallel) before it hits the grating or prism. After the dispersive element, the different wavelength components propagate in less different directions. The light may then pass through some additional optics and finally reach the photodetector. 
In order to obtain a spectrometer (non-scanning spectrometer), the photodetector can be a photodiode array, a CCD array or the like, recording the intensity in a certain spatial range, which corresponds to a certain spectral interval. One can obtain the spectrum directly by mapping the different wavelengths to the detector pixels. Since all wavelength components can be measured simultaneously, data acquisition is relatively fast. The resolution is typically limited by the pixel density of the detector, or may be limited by the optical setup. By using interpolation to determine the spectral peak position, it has better accuracy than according to the pixel pitch. 
In a scanning spectrometer, the detector may be a single photodiode or photomultiplier tube, placed behind a narrow optical slit so that only one narrow interval of wavelengths at a time can reach the detector. Then, assuming that the input PSD remains constant during that time, the slit position or angular orientation of the grating or prism can be moved so that a specific wavelength range can be scanned. The device then acts as a tunable monochromator. Figure 1 shows a common design of a Chernier-Turner grating monochromator. If a wide spectral range is scanned with high resolution, and if the detector is not very fast, the acquisition time of the full spectrum may be long, for example because low optical power must require a rather long averaging time. If the properties of the light source are not stable, the long acquisition time will not only cause inconvenience, but also become a troublesome problem. 

Figure 1: The design of the Chernier-Turner monochromator used in the scanning spectrometer, there is no exit slit (or wider slit) for non-scanning devices.

The light passing through the entrance slit is collimated by a curved mirror, undergoes a wavelength-dependent deflection at the diffraction grating, and is then focused again by another curved mirror. For one orientation of the diffraction grating, only light within a narrow wavelength band can pass through the exit slit. (The light shown is suitable for wavelengths within this interval) The entire device is placed in a box, which contains an aperture and a black housing to minimize the effect of stray light. 
There are also improved designs, such as the cross Chenier-Turner spectrometer, which allows for a more compact design than the unfolded version. Another version is based on concave holographic gratings and does not involve additional curved mirrors. 
Some grating spectrometers are very compact, only a few centimeters wide. However, the highest performance (especially in terms of resolution and sensitivity) is obtained with larger instruments. Typical wavelength resolutions achieved by such spectrometers are in the order of 0.01 nm to 0.1 nm. 
Depending on the type of spectrometer used (e. g. grating spectrometer), various details must be observed: 

• The input light typically has to be sent onto an entrance slit of variable width. In order to obtain the highest spectral resolution, the slit should be narrowed, but this reduces the emission power, so it may cause increased noise or longer acquisition time, especially for low-brightness light sources. Some spectrometers have fiber optic inputs, either multimode fiber or single mode fiber. Multimode fiber makes it easier to collect light, while single-mode fiber enables the highest spectrometer performance.
• Diffraction gratings typically use the first diffraction order, but sometimes higher diffraction orders are used for better performance. Regardless of the diffraction order, there may be problems with light and artifacts at other diffraction orders. If difficult-to-interpret spectral features are found, they can be examined for possible such artifacts.
• The response of the spectrometer may depend on the polarization, since the diffraction efficiency of the grating or the reflection loss on the prism arrangement is dependent on the polarization.
Some spectrometers must be calibrated by the user. For wavelength calibration, certain discharge lamps may be used, emitting a line spectrum with precisely defined wavelength components. The calibration of responsivity across the wavelength interval is often more difficult. One approach is to use an incandescent lamp with a known filament temperature or calibration spectrum.

interference spectrometer 
 
High spectral resolution can be obtained with various types of interference spectrometers, but only in a very limited spectral range: 

• Some instruments are based on a Fabry-Perot interferometer, in which the mirror spacing is scanned mechanically, for example by a piezoelectric linear actuator, while the transmitted optical power is recorded. The usable spectral interval is the so-called free spectral range, determined by the mirror distance; this is typically of the order of 0.1 GHz to 10 GHz, I .e. in nanometers. The resolution bandwidth is the free spectral range divided by the accuracy, which is mainly determined by the specular reflectance. A larger distance between the mirrors can achieve higher performance, but it also results in a narrowing of the free spectral range.
· The Fourier transform spectrometer contains a Michelson interferometer, in which one arm length can be mechanically scanned over a long distance (millimeters, centimeters, or even beyond). The time dependence of the detector signal recorded during the full arm length scan must be Fourier transformed to obtain the spectrum. The advantage is that there is a high spectral resolution and only a single photodetector is required.
A simplified version is the wavelength meter, which is specifically designed to accurately measure the wavelength of the laser source instead of recording the complete spectrum.
• Arrayed waveguide gratings for very compact spectrometers. They are based on interference effects in small waveguide devices.

Dispersion Spectrum Analysis Based on Propagation Time 

Completely different operations can be realized for the principle of spectrum analysis of broadband ultrashort pulses. One can simply send such a pulse through a long optical fiber, which introduces a large amount of dispersion and causes different spectral components to arrive at different times after the fiber: for example, a pulse with an original pulse duration far below 100 fs may be dispersed within a few nanoseconds. By analyzing this light using a photodiode and an oscilloscope, spectral information can be obtained. Of course, we should ensure that optical non-linear effects do not nullify the results; in optical fibers, this greatly limits the allowable peak power. 

A very attractive feature of this method is that spectra can be recorded at high speed. A single ultrashort pulse is sufficient for this purpose, whereas conventional scanning spectrometers may require many seconds and average the spectrum over the pulse train. 

Spectrometer for extreme spectral regions 

Conventional spectrometers operate in the visible spectral range or near infrared or ultraviolet light. However, there are also spectrometers that can operate in extreme spectral regions. 
Some devices work at short wavelengths, I .e. in the extreme ultraviolet (EUV) and even X-ray region, with wavelengths of only a few nanometers. Such devices exploit periodicity on an atomic scale and may consist of diffraction gratings with very small line spacing, gratings in the X-ray region or even single crystal gratings. As a photodetector, a special X-ray CCD camera or a microchannel plate (MCP) detector (→photomultiplier tube) can be used. 
Other spectrometers are suitable for the mid-infrared spectral region (→ mid-infrared spectrometers). They require special infrared optics and suitable detectors. Long wavelength photodetectors are available but have limited performance in terms of detection noise and bandwidth. Substantial improvements can be achieved by using a laser sum-frequency generation method to convert infrared light into visible or near-infrared light by an up-conversion effect in order to utilize visible or near-infrared photodetectors. This aspect is particularly important for spectrometers, where a multi-channel detector is required. Most of these devices use silicon technology and are only suitable for wavelengths below about 1 μm.

Application of Spectrometer

Because the application field of spectrometer is very wide. Some typical examples are as follows: 

In optical technology and fundamental physics, spectrometers are used to characterize various light sources and optical elements.
For example, optical telescopes in astronomy often use spectrometers to obtain additional information about galaxies, stars, and planets.
In chemistry, spectrometers can be used to identify substances or measure the concentration of multiple substances, such as in liquid solutions or gases. For more details, see the article on spectrophotometers. In the context of lidar, spectral analysis can be used for environmental monitoring.

See also the article on spectroscopy for more applications. 
Depending on the application, the requirements for the wavelength region covered, performance (e. g., resolution, sensitivity, speed, etc.), and cost may be very different. As a result, there are a wide variety of spectrometers on the market, and specialized versions have been developed for special applications, such as astronomy.